Model Number 71992-01 |
Device Problem
Device Displays Incorrect Message (2591)
|
Patient Problems
Hypoglycemia (1912); Discomfort (2330)
|
Event Date 08/09/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
The product has been requested back for an investigation.At this time product has not yet been returned.An extended investigation has been performed for the reported complaint and it has been determined that there was no indication that the product did not meet specification.The device history records (dhrs) for the fs libre sensor and fs libre sensor kit were reviewed and the dhrs showed the fs libre sensor and sensor kit passed all tests prior to release.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
An error message was reported with the abbott diabetes care (adc) device.Customer received a "replace sensor" message and was unable to obtain readings.As a result, customer experienced a loss of consciousness, "feeling unwell", and was unable to self-treat, requiring third-party treatment of "breakfast" by a healthcare professional (hcp) for a diagnosis of hypoglycemia.There was no report of death or permanent impairment associated with this event.
|
|
Manufacturer Narrative
|
Sensor (b)(6) was returned and investigated.The sensor plug was properly seated and no physical damage was observed on the sensor patch.Data was extracted from the returned sensor using approved software.The sensor was found to be in sensor state 5 (indicating a normal termination).Visual inspection was performed on the sensor plug and a broken sensor snaps was observed.An extended investigation has also been performed on the sensor plug and observed that the snaps were broken off, which may cause the sensor plug to be unable to form a proper seal and can lead to liquid ingress onto the pcb.Therefore, the issue is confirmed to a broken snap.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
An error message was reported with the abbott diabetes care (adc) device.Customer received a "replace sensor" message and was unable to obtain readings.As a result, customer experienced a loss of consciousness, "feeling unwell", and was unable to self-treat, requiring third-party treatment of "breakfast" by a healthcare professional (hcp) for a diagnosis of hypoglycemia.There was no report of death or permanent impairment associated with this event.
|
|