• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. 168 CM (66") APPX 2.4 ML, PUR YELLOW SMALLBORE EXT SET, 3-PORT NANOCLAVE¿ MANIFO; STOPCOCK, I.V. SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. 168 CM (66") APPX 2.4 ML, PUR YELLOW SMALLBORE EXT SET, 3-PORT NANOCLAVE¿ MANIFO; STOPCOCK, I.V. SET Back to Search Results
Catalog Number 011-AM3106
Device Problem Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/28/2023
Event Type  malfunction  
Event Description
It was reported that a168 cm (66") appx 2.4 ml, pur yellow smallbore ext set, 3-port nanoclave¿ manifold w/check valve, nanoclave¿, nanoclave¿ 4-way stopcock (red ring), rotating luer was found with the emergency line valve pressed, during infusion, and not usable for at least 48 hours.Additionally, the patient has an ¿arm catheter, placed centrally, since may 15,(b)(6) 2023.Iccu 8, changed on may 26, 2023.Continuous treatments: parenteral administration, fentanyl and hypnovel respectively at 6.5 ml/h; 0.3ml/h and 0.3ml/h.Discontinuous treatments: perfalgan and atb¿.There was no harm reported.
 
Manufacturer Narrative
The device is available to be returned for evaluation.However, it has not yet been received.
 
Manufacturer Narrative
The complaint was confirmed.The nanoclave on the nanoclave stopcock was observed to be stuck down and subsequent disassembly revealed a bent spike and torn seal.The probable cause of the seal sticking down is typical of access with an incompatible mating device during use.The dfu states: that needle free connectors are compatible with iso male luers having an internal diameter between 0.062 inches (1.58 mm) and 0.110 inches (2.8 mm).A device history review (dhr) could not be conducted because no lot number(s) was/were identified.Additional information: d9 - date returned to mfg - 9/14/2023.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
168 CM (66") APPX 2.4 ML, PUR YELLOW SMALLBORE EXT SET, 3-PORT NANOCLAVE¿ MANIFO
Type of Device
STOPCOCK, I.V. SET
Manufacturer (Section D)
ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V.
avenida cuarzo no. 250
ensenada, b.cfa. 22790
MX  22790
Manufacturer Contact
michael visocnik
600 n. field dr.
lake forest, IL 60045
2247062300
MDR Report Key17548934
MDR Text Key321182791
Report Number9617594-2023-00552
Device Sequence Number1
Product Code FMG
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K964435
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number011-AM3106
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/20/2023
Initial Date FDA Received08/15/2023
Supplement Dates Manufacturer Received10/19/2023
Supplement Dates FDA Received04/10/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
ARM CATHETER PLACED CENTRALLY MAY 15, 2023 UKN MFR; FENTANYL 0.3ML/H, UNK MFR; HYPNOVEL 0.3ML/H, UNK MFR; PARENTERAL ADMINISTRATION 6.5 ML/H, UNK MFR; PERFALGAN, UNK MFR; UNSPECIFIED ANTIBIOTICS (ATB), UNK MFR
Patient Age1 DA
Patient SexMale
Patient Weight1 KG
-
-