Model Number 011160-10
Device Problem Detachment of Device or Device Component (2907)
Patient Problem Insufficient Information (4580)
Event Date 07/27/2023
Event Type  Injury  
Manufacturer Narrative
The suspected device is pending return for evaluation.Once the device is returned and evaluation is completed, a supplemental report will be made.The event is filed under internal karl storz complaint id: (b)(4).
 
Event Description
It was reported that the end of the electrode bipolar cutting loop snapped off and went into the patient's bladder while the surgeon was carrying out a turp.The surgery was prolonged for approximately half an hour as the surgeon had to retrieve the broken part of the loop before continuing with the operation.The broken part was retrieved.
 
Manufacturer Narrative
Final investigation date 2023-09-06.Product was not returned for investigation.A definite root cause can't be determined.In similar cases the cutting loop got exposed to excessive force caused it to break.The event is filed under internal karl storz complaint id: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECTRODE, BIPOLAR
Type of Device
ENDOSCOPIC ELECTROSURGICAL ELECTRODE
Manufacturer (Section D)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM  78532
Manufacturer (Section G)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM   78532
Manufacturer Contact
anja fair
2151 e grand ave.
el segundo, CA 90245
4242188247
MDR Report Key17549124
MDR Text Key321164722
Report Number9610617-2023-00209
Device Sequence Number1
Product Code HIH
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K221893
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/15/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number011160-10
Device Catalogue Number011160-10
Device Lot NumberPM61
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received09/06/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/01/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
-
-