DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-3023-53 |
Device Problem
No Display/Image (1183)
|
Patient Problems
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/09/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that, the cs300 intra-aortic balloon pump (iabp) unit's display is blank.
|
|
Manufacturer Narrative
|
A supplemental report will be submitted upon completion of our investigation.
|
|
Event Description
|
It was reported that during a routine check , the cs300 intra-aortic balloon pump (iabp) unit's display is blank.
|
|
Manufacturer Narrative
|
A getinge fse evaluated the unit and provide the following details: on 8/24/2023, the fse verified which parts the unit needed and he was trying to duplicate the issue.On 8/25/2023, the fse did some diagnosing and the unit was pointing to main board or cable that leads to monitor but neither one of these parts corrected the issue according to the manual.The fse then did some troubleshooting and identified an issue with both datasets.The fse ordered the parts to replace them.On 9/8/2023, the fse replaced both datasets and the issue was resolved.The fse performed all functional and safety tests to factory specifications.The unit passed all tests performed and was returned to the customer.
|
|