• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIODYNAMICS NANOKNIFE SYSTEM; LOW ENERGY DIRECT CURRENT ABLATION SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIODYNAMICS NANOKNIFE SYSTEM; LOW ENERGY DIRECT CURRENT ABLATION SYSTEM Back to Search Results
Catalog Number 20400107
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Bradycardia (1751)
Event Date 07/26/2023
Event Type  Injury  
Manufacturer Narrative
As the reported defective device was not returned, since there was no reported nanoknife system malfunction during the procedure, angiodynamics is unable to perform a device evaluation.The customer's reported complaint description of patient cardiac issue (bradycardia) cannot be confirmed given the patient centric nature of this serious adverse event (sae).No nanoknife probe product was returned to angiodynamics for evaluation since there was no report of device or system imalfunction during the procedure, just patient sae post procedure.The bradycardia was corrected by placing the patient under general anesthesia and hospital monitoring.Heart arrhythmia (i.E.Bradycardia) is a potential adverse effect from a nanoknife system procedure and is cautioned against in the directions for use a review of the device history records was performed for the reported packaging and component lots for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly, and performance specifications.Labeling review: the instructions for use which is supplied to the end use, states: warnings: do not use a device with damaged insulation.Do not attach anything to the device unless it is supplied by angiodynamics and indicated for use with this device.Attachments may damage the insulation and contribute to patient injury.Adverse effects that may be associated with the use of the nanoknife system include, but are not limited to: arrhythmia, atrial, fibrillation or flutter, bigeminy, bradycardia, heart block or atrioventricular block, paroxysmal supraventricular tachycardia, tachycardia, reflex tachycardia, ventricular tachycardia, ventricular fibrillation, fistula formation, damage to critical anatomical structure (nerve, vessel, duct), hematoma, hemorrhage, hemothorax, infection, muscle contraction pneumothorax , reflex hypertension, unintended mechanical perforation, vagal stimulation, asystole and venous thrombosis.The user manual for the nanoknife generator, states: electrodes that are not parallel to each other may result in an incomplete ablation.Inappropriately positioned electrodes or metal implants in the field may distort the desired ablation field.Avoid unnecessarily high voltage or excessive number of pulses.Avoid short-circuiting the electrodes when delivering pulses.Electrode to electrode contact or electrode to electrode spacing less than 5 mm (millimeters) may result in short circuiting during energy delivery resulting in incomplete ablation.Adverse effects that may be associated with the use of the nanoknife system include, but are not limited to the following: arrhythmia, pneumothorax, muscle contraction, hemorrhage, unintended mechanical perforation, infection, bradycardia, vagal stimulation, asystole, damage to critical anatomical structure (nerve, vessel, and/or duct).".A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.(b)(4).
 
Event Description
Angiodynamics' senior clinical specialist reported that post patient's prostate ire procedure, the patient was in recovery and had to be placed under general anesthesia due to vital signs.The patient was reported to have experienced bradycardia approximately 15 to 20 minutes post procedure.The patient was transported in stable condition to the hospital for observation.There was no reported device malfunction or procedure issues that contributed to the patient's adverse event.It was reported that patient's co-morbidity was a possible contributing factor to the event.Per the nurse at the facility, the patient was improving.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NANOKNIFE SYSTEM
Type of Device
LOW ENERGY DIRECT CURRENT ABLATION SYSTEM
Manufacturer (Section D)
ANGIODYNAMICS
10 glens falls technical park
glens falls NY 12801
Manufacturer (Section G)
ANGIODYNAMICS
10 glens falls technical park
glens falls NY 12801
Manufacturer Contact
alexandra invencio
26 forest street
marlborough, MA 01752
5086587990
MDR Report Key17550653
MDR Text Key321225978
Report Number1317056-2023-00098
Device Sequence Number1
Product Code OAB
UDI-Device IdentifierH787204001070
UDI-PublicH787204001070
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K080287
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 08/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/15/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number20400107
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/26/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other; Hospitalization;
Patient Age63 YR
Patient SexMale
-
-