• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESMED LTD ASTRAL 150 - AMER; VENTILATOR, CONTINUOUS (FACILITY/HOME)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESMED LTD ASTRAL 150 - AMER; VENTILATOR, CONTINUOUS (FACILITY/HOME) Back to Search Results
Model Number 27003
Device Problems Electrical /Electronic Property Problem (1198); Protective Measures Problem (3015)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/17/2023
Event Type  malfunction  
Event Description
During evaluation by a third party service center, an astral device displayed an error message (sf179) related to the super capacitor.There was no patient harm or serious injury reported as a result of this incident.
 
Manufacturer Narrative
Visual inspection of the main circuit board revealed a leaky super capacitor.The main circuit board was replaced to address the reported complaint.The device was serviced and tested before it was returned to the customer.Based on all available evidence and complaint investigations of a similar nature, an investigation determined that the reported complaint was due to super capacitor electrolyte leak.Resmed's risk analysis for this failure mode concludes that the risk is acceptable.Resmed reference#: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASTRAL 150 - AMER
Type of Device
VENTILATOR, CONTINUOUS (FACILITY/HOME)
Manufacturer (Section D)
RESMED LTD
1 elizabeth macarthur drive
bella vista
sydney, nsw 2153
AS  2153
MDR Report Key17553223
MDR Text Key321285785
Report Number3007573469-2023-00472
Device Sequence Number1
Product Code NOU
UDI-Device Identifier00619498270033
UDI-Public(01)00619498270033(11)181211(10)1348343
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Distributor
Reporter Occupation 505
Type of Report Initial
Report Date 08/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/15/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number27003
Device Catalogue Number27003
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report Sent to FDA08/15/2023
Distributor Facility Aware Date07/17/2023
Device Age55 MO
Date Report to Manufacturer08/15/2023
Date Manufacturer Received07/17/2023
Date Device Manufactured12/16/2018
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-