WARSAW ORTHOPEDICS BRYAN CERVICAL DISC SYSTEM; PROSTHESIS, INTERVERTEBRAL DISC Back to Search Results
Model Number 6470316
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/19/2023
Event Type  Injury  
Manufacturer Narrative
Continuation of d10: section d information references the main component of the system.Other relevant device(s) are: product id: 6470316, serial/lot #: (b)(6), ubd: 10-may-2018, udi#: (b)(4) two bryan discs with lot 0247320w and 0263928w were implanted during initial surgery and it is unknown which device has failed.So, both of these devices are being reported and will be updated accordingly if additional information is received regarding the event on a later date.H6: the codes present in section h6 correspond to multiple components/products that comprise this reported event.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may have not been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from healthcare provider (hcp) via a manufacturer representative regarding a patient having c5/c6 cervical tdr for an indication of c5/c6 radiculopathy.It was reported that there was internal failure of cage not allowing to compress. two bryan discs with lot 0247320w and 0263928w were implanted, so it is not sure which lot number the failed cage came from.The bryan disc failed causing instability and the additional surgery was performed on (b)(6) 2023 to remove the implant and replaced with a corpectomy cage. there were no patient symptoms reported.There were no further complications reported regarding the event.
 
Manufacturer Narrative
Additional information: h10 radiographic image review: the intra-op view - cervical artificail disc placement, interior level difficult to visualise.Saggital ct post op - the superior interbody device appears to constructed poorly.Ap view of same - appears positioned correctly in coronal plane.Another saggital view is similar to the first one.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Additional information: d9, h3, h6 h3.Device evaluation summary: product analysis #(b)(4) : part# 6470316, lot# 0247320w.Visual inspection confirmed the disc has been used and there disc is coated in dried biological matter.Functional inspection confirmed the disc would not compress and return to its original position.Unable to determine root cause of issue due to the condition of the returned implant.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRYAN CERVICAL DISC SYSTEM
Type of Device
PROSTHESIS, INTERVERTEBRAL DISC
Manufacturer (Section D)
WARSAW ORTHOPEDICS
2500 silveus crossing
warsaw IN 46582
Manufacturer (Section G)
WARSAW ORTHOPEDICS
2500 silveus crossing
warsaw IN 46582
Manufacturer Contact
glen belmer
1800 pyramid place
memphis, TN 38132
6122713209
MDR Report Key17553749
MDR Text Key321222049
Report Number1030489-2023-00584
Device Sequence Number1
Product Code MJO
UDI-Device Identifier00613994181220
UDI-Public00613994181220
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P060023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 12/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/16/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/28/2018
Device Model Number6470316
Device Catalogue Number6470316
Device Lot Number0247320W
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/08/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/28/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
"SEE H10...."
Patient Outcome(s) Required Intervention;
Patient Age67 YR
Patient SexFemale
Patient Weight59 KG
-
-