• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. DREAMSTATION 2 ADV AUTO CPAP 2; VENTILATOR, NON-CONTINUOUS (RESPIRATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. DREAMSTATION 2 ADV AUTO CPAP 2; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number DSX520H11C
Device Problems Device Emits Odor (1425); Failure to Power Up (1476)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/16/2021
Event Type  malfunction  
Event Description
The manufacturer received information alleging the dreamstation 2 advanced auto cpap device fell off nightstand/water in unit.There was no report of patient harm or injury.The device was returned to the manufacturer's product investigation laboratory for evaluation.In evaluation it is find that the device is unit not powering on, and the pca had a thermal event due to water ingress.This mode of failure (water ingress) is covered under the current risk management file.Also noted a strong smell of burnt electronics.Both the pca and the touchscreen component attached to the ui bezel show thermal damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION 2 ADV AUTO CPAP 2
Type of Device
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15208
4125423300
MDR Report Key17556783
MDR Text Key321364595
Report Number2518422-2023-19874
Device Sequence Number1
Product Code BZD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K200480
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberDSX520H11C
Device Catalogue NumberDSX520H11C
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/21/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/16/2021
Initial Date FDA Received08/16/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/20/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-