Model Number CD3371-40QC
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/03/2023
Event Type  malfunction  
Event Description
It was reported that during an unrelated procedure, the set screw of the device was unable to be tightened.The device was explanted and replaced to resolve the event.The patient was in stable condition.
 
Manufacturer Narrative
The reported event of setscrew connection anomaly was confirmed.Analysis of the device revealed the atrial setscrew was stripped and contained septum material inside the hex cavity.This material in the hex cavity prevented full insertion of the torque driver and was the cause of the reported event.The set screw anomaly was consistent with having occurred during the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA MP ICD
Type of Device
NO MATCH
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17558214
MDR Text Key321273948
Report Number2017865-2023-38474
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/16/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2023
Device Model NumberCD3371-40QC
Device Lot NumberP000124013
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/19/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/02/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
BIOTRONIK RIGHT ATRIAL LEAD; BIOTRONIK RIGHT VENTRICULAR LEAD
-
-