Model Number DTMA1QQ
Device Problem Electromagnetic Interference (1194)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/26/2023
Event Type  malfunction  
Event Description
It was reported that electromagnetic interference (emi) sensed by the cardiac resynchronization therapy defibrillator (crt-d) resulted in the device triggering a lead integrity alert for high rate non-sustained (hrns) episodes and short ventricular intervals.Emi also caused a ventricular fibrillation (vf) episode to be detected; however, high voltage therapy was aborted.The emi was able to be observed on the hrns and vf episodes.The source of the emi was potentially thought to be an ungrounded swimming pool.The patient is to be seen for a device interrogation to disable the associated audible alert.The crt-d remains in use.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLARIA MRI QUAD CRT-D SURESCAN
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
SZ  1131
Manufacturer (Section G)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
SZ   1131
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key17560469
MDR Text Key321357021
Report Number9614453-2023-02835
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00763000059514
UDI-Public00763000059514
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 08/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/16/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date12/28/2021
Device Model NumberDTMA1QQ
Device Catalogue NumberDTMA1QQ
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/26/2023
Date Device Manufactured06/30/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
6935M62 LEAD, 407652 LEAD, 479888 LEAD
Patient Age47 YR
Patient SexMale
-
-