• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET; NO MATCH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET; NO MATCH Back to Search Results
Model Number 1458Q/92
Device Problem Failure to Advance (2524)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/04/2023
Event Type  malfunction  
Event Description
It was reported that the patient presented for a procedure.During the procedure, it was found that the stylet was failing to advance in the left ventricle (lv) lead.The lv lead was not used.The physician continued with another lv lead and completed the procedure.There were no patient consequences reported.
 
Manufacturer Narrative
The reported event of ¿stylet stuck in lead¿ was confirmed.A complete lead with stylet stuck was returned in one piece.The polytetrafluoroethylene (ptfe) coating of the stylet was stripped and was found bunched up with the inner coil at the connector region.The cause of the reported event was isolated to the bunching of the stylet ptfe coating inside the inner coil at the connector region that prevented the removal of the stylet consistent with damage cause during the procedure.
 
Manufacturer Narrative
Correction: the product was received.Therefore, the box "not returned to manufacturer" should not have been checked in h3.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17562309
MDR Text Key321317948
Report Number2017865-2023-38547
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734503358
UDI-Public05414734503358
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 08/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2023
Device Model Number1458Q/92
Device Lot NumberA000100758
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/08/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/04/2023
Initial Date FDA Received08/17/2023
Supplement Dates Manufacturer Received08/23/2023
08/24/2023
Supplement Dates FDA Received08/23/2023
08/28/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/05/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age81 YR
Patient SexFemale
Patient Weight73 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-