ABBOTT GALLANT HF; NO MATCH Back to Search Results
Model Number CDHFA500Q
Device Problems Defective Device (2588); Connection Problem (2900)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/02/2023
Event Type  malfunction  
Event Description
During the initial implant procedure, the atrial lead was unable to be connected to the header of the device.Upon investigation, a loose spring was noted in the atrial port of the device.A replacement device was used to complete the implant procedure.The patient was in stable condition.
 
Manufacturer Narrative
The reported field event of header anomaly and lead connection issue was confirmed.Visual analysis shows the atrial ring spring was dislodged and pushed into the atrial lead bore.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALLANT HF
Type of Device
NO MATCH
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17564273
MDR Text Key321337964
Report Number2017865-2023-38551
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05415067032010
UDI-Public05415067032010
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCDHFA500Q
Device Lot NumberS000088476
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/02/2023
Initial Date FDA Received08/17/2023
Supplement Dates Manufacturer Received11/14/2023
Supplement Dates FDA Received11/22/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/02/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-