OLYMPUS WINTER & IBE GMBH TELESCOPE, 10 MM, 0°, HD, QUICK LOCK, AUTOCLAVABLE; LAPAROSCOPY SYSTEM
|
Back to Search Results |
|
Model Number WA53000A |
Device Problem
Component Missing (2306)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/24/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device was returned and evaluated, and the customer¿s allegations (light guide missing and uneven lighting) were not confirmed.The investigation is ongoing, and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided.
|
|
Event Description
|
The customer reported that the front-end light guide of a telescope, 10 mm, 0°, hd, quick lock, autoclavable was missing.The event occurred during receipt inspection.There was no patient harm associated with the event.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely that the event occurred due to use of excessive force.Olympus will continue to monitor field performance for this device.
|
|