• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MERLIN PROGRAMMER; No Match

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MERLIN PROGRAMMER; No Match Back to Search Results
Model Number 3650
Device Problems Failure to Power Up (1476); Sparking (2595); Noise, Audible (3273)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The reported event of programmer unable to turn on and audible noise and visual sparking from programmer power source were not confirmed by analysis.The unit was plugged in to an external power source and powered on.The system was able to boot up successfully.During analysis, a failure event was observed which was unrelated to the reported event.Final analysis found damage to the printer.
 
Event Description
It was reported that the programmer was unable to be turned on.Audible noise and visual sparking was observed from the power source of the programming.No adverse consequences were reported for the patient or the user.The programmer was replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERLIN PROGRAMMER
Type of Device
No Match
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17568314
MDR Text Key321400606
Report Number2017865-2023-38558
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734500661
UDI-Public05414734500661
Combination Product (y/n)N
Reporter Country CodeEZ
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 08/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/17/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3650
Was the Report Sent to FDA? No
Date Manufacturer Received07/24/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/02/2010
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-