This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and a correction to d5.The device history record (dhr) was unable to be reviewed for this device since the lot was not provided, however the dhr for over the past year prior to the date of occurrence was inspected and the review found no deviations that could have caused or contributed to the reported issue.Olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Although a definitive root cause cannot be identified of the broken probe, the following step-by-step scenario likely caused the event: 1.When seal & cut output at pre-use inspection, the probe came in contact with a metal or a surgical instrument.Consequently, a scratch was made on the probe.2.The probe received an output load in seal & cut mode.As a result, the probe cracked from the scratch.3.The probe broke when added load in pre-use inspection.The following is included in the instructions for use: "do not grasp or let the probe tip contact hard objects such as metal clips, stapler, or other instruments (e.G., uterine manipulator).Also, be careful to avoid contacting the probe tip with those accidentally.Particularly during activation, a scratch on the probe tip could occur due to ultrasonic vibration, which leads the probe tip to break and fall off into the body cavity.In addition, the high-frequency (rf bipolar) current flows through the metal and generates spark discharge, which may cause burns and decrease functionalities".Olympus will continue to monitor field performance for this device.
|