This will be filed to damage to the soft tip.It was reported that a mitraclip procedure was performed to treat functional mitral regurgitation (mr) grade 4+ with an anterior leaflet flail.The physician advanced the clip into the ventricle (lv), but grasping was noted to be difficult.The clip was repositioned, but it was observed the posterior gripper would not lower.Troubleshooting was performed, but the issue was unable to be resolved.The physician then attempted to retract the clip back into the left atrium (la).However, resistance was felt while retracting the delivery catheter (dc) handle.The clip was able to be pulled back into the la, but at this time, a pericardial effusion was observed and the patient's blood pressure decreased, resulting in a clinically significant delay in the procedure.Pericardiocentesis was performed, but the effusion and blood pressure did not change.The clip was then attempted to be closed in an attempt to retract, but the clip was unable to close.The physician decided to retract the clip into the steerable guide catheter (sgc).However, the clip became caught on the soft tip of the sgc, causing damage to the soft tip and a bend in one of the clip arms.Both the sgc and clip delivery system (cds) were attempted to be removed together, but became stuck in the vein.At this point, the patient went into cardiac arrest and compressions were administered.The mitraclip devices were then forcefully removed from the patient in order to perform open heart surgery.It was noted the forceful removal of the device caused the clip to detach from the mandrel.Open heart surgery was attempted, but was unsuccessful as the patient bleed out and died.No additional information was provided.
|
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar complaints reported from this lot.Based on available information, the reported sgc deformation (soft tip) was due to procedural circumstances as retracting an opened clip into the sgc would likely result in sgc soft tip damage.There is no indication of a product quality issue with respect to manufacture, design, or labeling.The additional mitraclip device referenced in b5 is being filed under a separate medwatch report number.
|