• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PKG, CROSSFIRE 2 CONSOLE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PKG, CROSSFIRE 2 CONSOLE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Catalog Number 0475100000
Device Problems Electrical Shorting (2926); Activation Failure (3270)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/08/2023
Event Type  malfunction  
Event Description
Per investigation results, the device had blown components resulting in the potential for a thermal event.
 
Manufacturer Narrative
Alleged failure: it was reported that the crossfire 2 displayed an e41 error code immediately upon being opened and plugged in.This was an out of box failure.The failure(s) identified in the investigation is consistent with the complaint record.The root cause is the power board.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG, CROSSFIRE 2 CONSOLE
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
lucas wolski
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key17575882
MDR Text Key321514004
Report Number0002936485-2023-00838
Device Sequence Number1
Product Code GEI
UDI-Device Identifier07613327058109
UDI-Public07613327058109
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K071859
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 08/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number0475100000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/23/2023
Initial Date Manufacturer Received 07/27/2023
Initial Date FDA Received08/18/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-