Catalog Number 0475100000 |
Device Problems
Electrical Shorting (2926); Activation Failure (3270)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/08/2023 |
Event Type
malfunction
|
Event Description
|
Per investigation results, the device had blown components resulting in the potential for a thermal event.
|
|
Manufacturer Narrative
|
Alleged failure: it was reported that the crossfire 2 displayed an e41 error code immediately upon being opened and plugged in.This was an out of box failure.The failure(s) identified in the investigation is consistent with the complaint record.The root cause is the power board.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
|
|