Model Number SONNET CHARGING UNIT [OPUS 3 CHARGER] |
Device Problem
Overheating of Device (1437)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
A sonnet charging unit was received at service and repair with a burned usb port.
|
|
Manufacturer Narrative
|
The device has been returned to med-el hq where it will be evaluated.When available, a device failure analysis will be submitted as a follow-up report.
|
|
Manufacturer Narrative
|
Conclusion: a sonnet charging unit was sent to service _ repair because of malfunction.During device investigation a burned socket was detected which was clearly the reason for the reported issue.Electrical inspection could not be performed because of the observed failure.It must be assumed that the mentioned behaviour was caused by a bad contact in the socket of the charger due to a conductive substance e.G.Sweat/dirt.This is a final report.
|
|
Event Description
|
A sonnet charging unit was received at service and repair with a burned usb port which reportedly came back from a user.There was no report of an injury.
|
|