MEDTRONIC PUERTO RICO OPERATIONS CO. CLARIA MRI QUAD CRT-D SURESCAN; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
|
Back to Search Results |
|
Model Number DTMA1QQ |
Device Problems
Failure to Deliver Shock/Stimulation (1133); Pacing Problem (1439); Device Sensing Problem (2917)
|
Patient Problems
Dyspnea (1816); Tachycardia (2095)
|
Event Date 05/30/2023 |
Event Type
Injury
|
Event Description
|
It was reported that the right atrial (ra) lead exhibited loss of capture and undersensing.It was also noted that a loss of biventricular capture was observed.The leads remain in use.It was also reported that the cardiac resynchronization therapy defibrillator (crt-d) initially withheld therapy inappropriately for a ventricular tachycardia (vt) episode.It was also reported that the patient experienced shortness of breath.The crt-d remains in use.No further patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the right atrial (ra) lead exhibited loss of capture and undersensing.It was also noted that a loss of biventricular capture was observed on the left ventricular (lv) lead.The leads remain in use.It was also reported that the cardiac resynchronization therapy defibrillator (crt-d) initially withheld therapy inappropriately for a ventricular tachycardia (vt) episode.It was also noted that the device may have delivered inappropriate anti-tachycardia pacing (atp) for a supra ventricular tachycardia (svt) episode detected as vt.It was also reported that the patient experienced shortness of breath.The crt-d remains in use.No further patient complications have been reported as a result of this event.
|
|