Catalog Number UNKNOWN |
Device Problem
Fracture (1260)
|
Patient Problems
Failure of Implant (1924); Pain (1994)
|
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).Multiple mdr reports were filed for this event, please see associated reports: 0001825034 -2023 -01955.G2: foreign: netherlands.It is unknown if product will be returning to zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up report will be submitted.H3 other text : product not returned.
|
|
Event Description
|
It was reported that the patient was indicated for a revision due to implant fracture and pain approximately ten years post implantation.No additional patient consequences were reported.Attempts have been made and additional information on the reported event is unavailable at this time.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: three views of the right shoulder demonstrate a reverse total shoulder arthroplasty.Image three demonstrates intact hardware.Image one and two demonstrate fractured humeral hardware at the level of the tray.A fracture of the glenoid baseplate is not definitely confirmed.However, there may be suggestion of a baseplate fracture.Device history record (dhr) review was unable to be performed as the lot number of the device involved in the event is unknown.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.The reported event for the tray is confirmed through radiographs.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.D10: comp rvrs shdr glen bsplt +ha cat: 118001 lot: 721790.Versa-dial/comp ti std taper cat: 115330 lot: 591400.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|