Model Number VH-4000
Device Problem Failure to Deliver Energy (1211)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/25/2023
Event Type  malfunction  
Event Description
The hospital reported that during an endoscopic vein harvesting procedure, vasoview hemopro 2 tip looked frayed, did not cut or cauterize well.Not delivering energy.Used new device.No delay.No apparent harm to patient.
 
Manufacturer Narrative
Tw id# (b)(4).The device has been returned to the factory and is being evaluated.A supplemental report will be submitted when the evaluation is completed.
 
Event Description
N/a.
 
Manufacturer Narrative
Trackwise id# (b)(4).Specific actions for the reported/analyzed failure mode are being maintained and documented under maquet's corrective and preventive action (capa) system.The device was returned to the factory for evaluation on 08/08/2023.An investigation was conducted on 08/15/2023.A visual inspection was conducted.Signs of clinical use and evidence of charred material was observed on the heater wire.The heater wire was observed to be flexed away from the hot jaw from the base of the hot jaw with detachment at the tip of the hot jaw.There were no visual defects observed on the intact clear silicone insulation on both the cold and hot jaws.An electrical evaluation was conducted.A pre-cautery test was performed per the direction for use (dfu) with a reference cable, adapter, and reference power supply vh-3010 at level 3.0.The device passed the pre-cautery test; it produced visible steam and heat during ten (10) 3-second activations and shut off when the toggle was released.To evaluate the safety shut down system, a polyfuse activation test was performed 5 times over 10 minutes.The device shut off after the period of sustained activation and reactivated after 10-second cooling period with no incident each time.An activation and transection capability test was performed over four (4) repetitions using "max life test method stm2048073 rev aa.The device successfully transected tissue four (4) times.The jaws were gently cleaned of debris and char with a saline and gauze pad as indicated in the direction for use (cv000008979).No final testing was conducted due to the condition of the heater wire.Based on the results of the evaluation, the reported failures "failure to deliver energy¿ was not confirmed, however the analyzed failure "material twisted/ bent wire" was observed.The lot # 3000317188 history record review was completed.There were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VASOVIEW HEMOPRO 2
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer (Section G)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 
MDR Report Key17592713
MDR Text Key321677006
Report Number2242352-2023-00692
Device Sequence Number1
Product Code GEI
UDI-Device Identifier00607567700406
UDI-Public00607567700406
Combination Product (y/n)N
PMA/PMN Number
K101274
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 09/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/21/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberVH-4000
Device Catalogue NumberVH-4000
Device Lot Number3000317188
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/08/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/07/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/08/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Age68 YR
Patient SexMale
-
-