Model Number CENTRAL POST/SCREW TREPHINE
Device Problem Device Damaged by Another Device (2915)
Patient Problem Failure of Implant (1924)
Event Date 07/24/2023
Event Type  malfunction  
Event Description
On (b)(6) 2023, it was reported by a sales representative via email that the trephine reamer welded the post within it.It could not be taken apart, damaging the central post/screw trephine adapter as well.This was discovered when removing the central post after the base plate was removed during a revision rtsa procedure.Additional information received on 7/26/2023: there is no known information regarding the original procedure other than it was performed in (b)(6).The patient complained of pain since the original procedure was performed.Because of this, the patient underwent revision surgery on (b)(6) 2023.
 
Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRAL POST/SCREW TREPHINE
Type of Device
ORTHOPEDIC MANUAL SURG INSTR
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
emily shafer
8009337001
MDR Report Key17595737
MDR Text Key321727803
Report Number1220246-2023-07580
Device Sequence Number1
Product Code LXH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date 08/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/22/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCENTRAL POST/SCREW TREPHINE
Device Catalogue NumberAR-9662-R
Was Device Available for Evaluation? No
Date Manufacturer Received07/24/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-