|
Model Number ANGLED REAMER SLEEVE, 10° |
Device Problem
Device Damaged by Another Device (2915)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/20/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The complaint allegation was confirmed.One unpackaged ar-9597-10 serial/batch number (b)(6) was received for investigation.Functional testing with the returned ar-9676 batch 02210 and 022244 reveals the devices fit together.However, the devices do not rotate smoothly.Visual evaluation observed scratches lines around the outside diameter close to the proximal and distal ends.The most likely reason for the reported failure is the wear and tear damage incurred over repeated usage.Per dfu-0023 at revision 3.C.Validation.Repeated processing has minimal effect on these devices.End of life is normally determined by wear and damage due to the intended use.
|
|
Event Description
|
On (b)(6) 2023, it was reported by a sales representative via sems that an ar-9597-10 angled reamer sleeve heat welded to an ar-9676 angled reamer, along with a second ar-9676 angled reamer was damaged.This occurred during a revers total shoulder case on (b)(6) 2023, when reaming the glenoid with the mgs augment reamer, the surgeon noticed that it was not performing the way it typically does.After removing the reamer and handing it off to the back table for the tech to check it, they noticed that you could not disconnect the driver shaft and outer sleeve because they had heat welded together.They opened another mgs augment tray to complete the case, but when trying to separate the drive shaft from that 10-degree augment sleeve the driver shaft was damaged.There was no patient effect reported.
|
|
Search Alerts/Recalls
|
|
|