Model Number 6662 |
Device Problems
Battery Problem (2885); No Apparent Adverse Event (3189)
|
Patient Problems
Failure of Implant (1924); Inadequate Pain Relief (2388); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/02/2023 |
Event Type
Injury
|
Event Description
|
It was reported that the patient's ipg was nearing end of life but still operable.Surgical intervention was undertaken on (b)(6) 2023 wherein the patient's ipg was explanted, replaced, and therapy was restored.
|
|
Manufacturer Narrative
|
Section b3: date of event is estimated.
|
|
Manufacturer Narrative
|
Correction: manufacturing reference number should not have been reported as a medical device report (mdr).The reported observation of normal ipg battery depletion was confirmed.The root cause of the reported observation was due to the device having normal battery depletion at the time of the observation.The estimated longevity would have been 1.75 years +/-.25-year per ¿ 90205144, when using the as received programmed settings and assuming 500-ohm program impedances, for device model 6662.The total stimulation on time was 1.51 years, suggesting the device had normal battery depletion at the time of the observation.
|
|
Event Description
|
Additional information received indicated that product analysis findings confirmed the ipg exhibited normal battery depletion.
|
|
Search Alerts/Recalls
|