Model Number 71992-01 |
Device Problem
Detachment of Device or Device Component (2907)
|
Patient Problem
Device Embedded In Tissue or Plaque (3165)
|
Event Date 08/01/2023 |
Event Type
Injury
|
Event Description
|
A hcp reported the filament of the adc device remained in customer's skin.The customer was "vomiting blood" and was going to the hospital in an ambulance when a healthcare professional was removing the sensor and found the tip had remained inside the customer's arm upon removal.The sensor filament was removed by a nurse and no further treatment was reported.There was no report of death or permanent impairment associated with this event.
|
|
Manufacturer Narrative
|
The product has been requested back for an investigation.At this time product has not yet been returned.An extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.The device history records (dhrs) for the freestyle libre sensor and freestyle libre sensor kit were reviewed and the dhrs showed the freestyle libre sensor and sensor kit passed all tests prior to release.If the product is returned, a physical investigation will be performed and a follow-up report submitted.The date of event is unknown.The date entered in section b3 is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
|
|
Manufacturer Narrative
|
Sensor (b)(6)has been returned and investigated.Visual inspection has been performed on the returned sensor and no issues were observed.The sensor plug was properly seated.Severed sensor tail was observed.Visual inspection has been performed on the sensor plug assembly, no failure modes were observed.No further investigation performed as no product was returned.Issue closed to no product returned.If partial product is returned, the case will be re-opened, and a physical investigation will be performed.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
A hcp reported the filament of the adc device remained in customer's skin.The customer was "vomiting blood" and was going to the hospital in an ambulance when a healthcare professional was removing the sensor and found the tip had remained inside the customer's arm upon removal.The sensor filament was removed by a nurse and no further treatment was reported.There was no report of death or permanent impairment associated with this event.
|
|