Model Number UM-S20-17S
Device Problem Use of Device Problem (1670)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/20/2023
Event Type  malfunction  
Event Description
The customer reported to olympus that during a procedure the ultrasonic probe had defects.The procedure was completed using a similar device.There was no patient harm associated with the event.The device was evaluated, and it was found that the ultrasound probe could not be used.This mdr (medical device report) is being submitted to capture the reportable malfunction found during the device evaluation.
 
Manufacturer Narrative
The device was returned and evaluated, and the customer¿s allegation was confirmed.In addition to the ultrasonic probe not able be used, additional findings were as follows: the transparent part at the tip of the insertion tube was stretched and ultrasound image was not output.The investigation is ongoing, and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the event could not be determined although it can be presumed it was due that the ultrasonic medium inside the sheath leaked from the rupture point because the insertion part sheath reached the rupture.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRASONIC PROBE
Type of Device
ULTRASONIC PROBE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17611014
MDR Text Key321821181
Report Number3002808148-2023-08866
Device Sequence Number1
Product Code ITX
UDI-Device Identifier04953170368479
UDI-Public04953170368479
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K982323
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 11/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUM-S20-17S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/04/2023
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/03/2023
Initial Date FDA Received08/24/2023
Supplement Dates Manufacturer Received11/27/2023
Supplement Dates FDA Received11/29/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/01/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-