ABBOTT QUADRA ASSURA MP ICD; NO MATCH Back to Search Results
Model Number CD3371-40QC
Device Problems High impedance (1291); Loose or Intermittent Connection (1371)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/04/2023
Event Type  malfunction  
Event Description
It was reported that during the implant procedure, high pacing impedance was noted on the device due to a loose set screw.The set screw was unable to be tightened.The device was explanted and replaced to resolve the event.The patient was in stable condition.
 
Manufacturer Narrative
The reported event of set screw unable to be tightened and high pacing impedance could not be confirmed.The device was above the elective replacement indicator (eri) upon receipt.The ventricular set screw was not returned, so no analysis could be performed on it.A laboratory set screw was able to engage the ventricular connector block normally, and it was able to secure and release a test lead normally.The set screw anomaly was consistent with having occurred during the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA MP ICD
Type of Device
NO MATCH
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17617178
MDR Text Key321886033
Report Number2017865-2023-39234
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeTU
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 03/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/24/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCD3371-40QC
Device Lot NumberP000163934
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/01/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/23/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR
-
-