• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION JETSTREAM XC ATHERECTOMY CATHETER; CATHETER, PERIPHERAL, ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION JETSTREAM XC ATHERECTOMY CATHETER; CATHETER, PERIPHERAL, ATHERECTOMY Back to Search Results
Model Number 45007
Device Problems Detachment of Device or Device Component (2907); Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/01/2023
Event Type  malfunction  
Manufacturer Narrative
E1 - initial reporter phone: (b)(6).
 
Event Description
It was reported that the guidewire was unable to pass though the jetstream.This 2.1mm jetstream xc atherectomy catheter was selected for use for this atherectomy procedure of the femoral-popliteal with severe tortuosity and mild calcification.During the procedure, an attempt was made to advance the thruway guide and it did not pass due to a kink in the catheter.When the catheter was removed from the patient it was noticed that there was a fracture approximately 20 cm from the beginning of the catheter.The catheter was able to be removed in one piece.The system was flushed twice, and the procedure was completed successfully using another jetstream without sequalae.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JETSTREAM XC ATHERECTOMY CATHETER
Type of Device
CATHETER, PERIPHERAL, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
model farm road
cork T12 Y K88
EI   T12 YK88
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key17618893
MDR Text Key321918965
Report Number2124215-2023-45826
Device Sequence Number1
Product Code MCW
Combination Product (y/n)N
Reporter Country CodeCO
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 08/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/24/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number45007
Device Catalogue Number45007
Device Lot Number0031802160
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/01/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/12/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient SexFemale
-
-