Catalog Number UNK LIFESTENT VASCSTNT SYS |
Device Problem
Fracture (1260)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/27/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
H10: as the lot number for the device was not provided, a review of the device history records could not be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
|
|
Event Description
|
It was reported that approximately three months post stent placement, one of the stents allegedly fractured.There was no reported patient injury.
|
|
Manufacturer Narrative
|
H10: manufacturing review: as the lot number for the device was not provided, a review of the device history records could not be performed.Investigation summary: the sample was not returned for evaluation and no images were provided for review which leads to inconclusive evaluation result.A 6f introducer was in use, and the lesion was pre dilated; the stents deployed properly, and the related vessels were radiographically visualized well leading to a smooth procedure; no abnormality was found.It was not known whether the stents were placed overlapping, and where exactly the different stents were placed; it was even not known which of the three stents was fractured and there was one competitor product involved.A closer description of the fracture was not available, and the user did not provide images for verification.Based on the investigation of the provided poor information, the investigation is closed as inconclusive.A definite root cause for the reported event could not be determined.Labeling review: in reviewing the relevant labeling for this product the potential issue was found addressed.The instructions for use closely describes holding and handling of the system during deployment, in particular the instructions for use state: 'do not hold the silver stent delivery sheath at any time during deployment.Do not constrict the stent delivery sheath during stent deployment.' regarding pta the instructions for use state: 'predilation of the lesion should be performed using standard techniques.', and 'post stent expansion with a pta catheter is recommended.' regarding access / accessories the instructions for use state: 'gain femoral access utilizing a 6f (2.0 mm) or larger introducer sheath.Insert a 0.035 inch (0.89 mm) diameter guidewire'.The instructions for use further state: 'the safety and effectiveness of stent overlapping in the middle (p2) and distal popliteal artery (p3) has not yet been established.', and 'cases of fracture have been reported in clinical use of the lifestent vascular stent.Cases of stent fracture occurred in lesions that were moderate to severely calcified, proximal or distal to an area of stent overlap and in cases where stents experienced >10% elongation at deployment.Therefore, care should be taken when deploying the stent as manipulation of the delivery system may, in rare instances, lead to stent elongation and subsequent stent fracture.The long-term clinical implications of these stent fractures have not yet been established.' a stent size selection table is part of the instructions for use describing the relationship between reference vessel diameter and stent diameter.H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
Event Description
|
It was reported that approximately three months post stent placement, the stent was allegedly fractured.There was no reported patient injury.
|
|