MAUDE Adverse Event Report: ANGIOMED GMBH & CO. MEDIZINTECHNIK KG LIFESTENT VASCULAR STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIOMED GMBH & CO. MEDIZINTECHNIK KG LIFESTENT VASCULAR STENT Back to Search Results
Catalog Number UNK LIFESTENT VASCSTNT SYS
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/27/2021
Event Type  malfunction  
Manufacturer Narrative
H10: as the lot number for the device was not provided, a review of the device history records could not be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Event Description
It was reported that approximately three months post stent placement, one of the stents allegedly fractured.There was no reported patient injury.
 
Manufacturer Narrative
H10: manufacturing review: as the lot number for the device was not provided, a review of the device history records could not be performed.Investigation summary: the sample was not returned for evaluation and no images were provided for review which leads to inconclusive evaluation result.A 6f introducer was in use, and the lesion was pre dilated; the stents deployed properly, and the related vessels were radiographically visualized well leading to a smooth procedure; no abnormality was found.It was not known whether the stents were placed overlapping, and where exactly the different stents were placed; it was even not known which of the three stents was fractured and there was one competitor product involved.A closer description of the fracture was not available, and the user did not provide images for verification.Based on the investigation of the provided poor information, the investigation is closed as inconclusive.A definite root cause for the reported event could not be determined.Labeling review: in reviewing the relevant labeling for this product the potential issue was found addressed.The instructions for use closely describes holding and handling of the system during deployment, in particular the instructions for use state: 'do not hold the silver stent delivery sheath at any time during deployment.Do not constrict the stent delivery sheath during stent deployment.' regarding pta the instructions for use state: 'predilation of the lesion should be performed using standard techniques.', and 'post stent expansion with a pta catheter is recommended.' regarding access / accessories the instructions for use state: 'gain femoral access utilizing a 6f (2.0 mm) or larger introducer sheath.Insert a 0.035 inch (0.89 mm) diameter guidewire'.The instructions for use further state: 'the safety and effectiveness of stent overlapping in the middle (p2) and distal popliteal artery (p3) has not yet been established.', and 'cases of fracture have been reported in clinical use of the lifestent vascular stent.Cases of stent fracture occurred in lesions that were moderate to severely calcified, proximal or distal to an area of stent overlap and in cases where stents experienced >10% elongation at deployment.Therefore, care should be taken when deploying the stent as manipulation of the delivery system may, in rare instances, lead to stent elongation and subsequent stent fracture.The long-term clinical implications of these stent fractures have not yet been established.' a stent size selection table is part of the instructions for use describing the relationship between reference vessel diameter and stent diameter.H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that approximately three months post stent placement, the stent was allegedly fractured.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFESTENT VASCULAR STENT
Type of Device
VASCULAR STENT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM  76227
Manufacturer (Section G)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM   76227
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key17622192
MDR Text Key321932462
Report Number9681442-2023-00304
Device Sequence Number1
Product Code NIP
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P070014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/25/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNK LIFESTENT VASCSTNT SYS
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/01/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-