DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-3023-53 |
Device Problem
Premature Discharge of Battery (1057)
|
Patient Problems
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/15/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
A supplemental report will be submitted upon completion of our investigation.
|
|
Event Description
|
It was reported that, the cs300 intra-aortic balloon pump (iabp) unit's battery is not lasting more than 10 min.
|
|
Manufacturer Narrative
|
Corrected data: b5, h6 (clinical and impact code).Updated data: b4, g3, g6, h2, h10, h11.
|
|
Event Description
|
It was reported that during use on a patient, the cs300 intra-aortic balloon pump (iabp) unit's battery is not lasting more than 10 min.While walking the patient the low battery alarm sounded so the patient was taken back to their room and plugged the device in.Therapy never stopped.There was no patient harm.
|
|
Manufacturer Narrative
|
A getinge field service engineer (fse) evaluated the unit and replaced the batteries due to short run time.The fse then completed all functional and safety tests to factory specifications.The unit was returned to the customer.
|
|
Search Alerts/Recalls
|
|
|