Catalog Number EL5ML
Device Problems Mechanical Problem (1384); Failure to Form Staple (2579); Mechanical Jam (2983)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/08/2023
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Date sent: 8/25/2023.D4: batch # a9c86n.Additional information was requested and the following was obtained: "did device not feed clips? ( yes).Did device feed clips sideways?( yes).Did device not fire clips (jammed)? (no, device fired).Did device fire malformed clips? (yes).Did device fire scissored clips? (yes).Did device drop or eject clips? (yes).If other, please specify".Investigation summary.The product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the el5ml device was returned with no apparent damage.In an attempt to replicate the reported incident, the instrument was tested for functionality.During the analysis, the device was cycled and it fed and formed 11 conforming clips.Upon testing, the jaws open and close without any difficulties.In addition, the device locked out as intended.  although no conclusion could be reached on the cause of the reported event, the instructions for use do contain the following caution: do not excessively twist or torque the instrument jaws when positioning or firing the instrument on a tubular structure or vessel.Excessive twisting or torquing may result in clip malformation.Do not insert the clip applier through a trocar if a clip is present in the jaws.This may result in clip malformation, dislodged clips, or damage to the instrument.If a clip is present in the jaws, fully squeeze the trigger against the handle, then fully release the trigger to release the clip from the jaws before inserting the device through the trocar.The event described could not be confirmed as the device performed without any difficulties noted.As part of ethicon¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.A manufacturing record evaluation was performed for the finished device batch and lot number, and no non-conformance were identified.
 
Event Description
It was reported that during a laparoscopic cholecystectomy, the clips did not fire properly.Procedure successfully completed.No patient consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGAMAX-5MM ENDO CLIP APPLIER
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
*  
3035526892
MDR Report Key17628332
MDR Text Key322038532
Report Number3005075853-2023-06106
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10705036001843
UDI-Public10705036001843
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K050344
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other
Type of Report Initial
Report Date 08/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/25/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberEL5ML
Device Lot NumberA9C86N
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/24/2023
Date Manufacturer Received07/29/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/10/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-