There are controls in the manufacturing process to ensure the product met specifications upon release.The subject device was returned for analysis.The guidewire was noted to be kinked at 79cm from the proximal end of the guidewire.The guidewire nitinol tubing was noted to be broken/fractured at 87cm from the proximal end.The proximal and distal fragment was inspected, and the nitinol tubing was broken/fractured with the core wire and platinum wire still connected but stretched and broken.The guidewire ptfe (polytetrafluoroethylene) coating was seen to be peeling at 105cm and 121cm from the proximal end of the guidewire.The core wire distal end was observed through the distal tip dome.The guidewire distal tip was hydrating, no anomalies were noted.Functional inspection: guidewire difficult to advance functional test could not be performed due to damage (fracture).The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported events can be confirmed based on analysis.The device failed to meet specifications when received for complaint investigation based on the analyzed anomalies noted to the device.It was reported that during delivery the operator felt friction.Withdrew the guidewire out and found the coating of it peeled off and kinked.Additional information provided by the customer indicated that the device was prepared for use as per the directions for use, the device was confirmed to be in good condition during preparation/prior to use on the patient, continuous flush was set up and maintained throughout the clinical procedure and the patients anatomy was severely tortuous.During analysis it was noted that the guidewire had been broken/fractured, stretched, the guidewire was kinked and there was peeling noted to the ptfe coating.It is probable that the guidewire experienced tension and/or torsion forces while navigating inside the anatomy of the patient, due to the tortuous anatomy of the patient resulting in the kinking, fracture and stretching noted to the guidewire.An assignable cause of procedural factors will be assigned to the reported guidewire difficult to advance and guidewire kinked/bent and to the analyzed guidewire kinked/bent, guidewire distal tip stretched and guidewire broken/ fractured during use, as the issue is associated with a product that meets stryker design and manufacture specifications and was used in according with the dfu but due to procedural and/or anatomical factors during use, the product performance was limited.The peeling noted to the ptfe coating is indicative of interaction of the wire while backloading of the wire through the introducer.The as analyzed ¿guidewire ptfe coating peeling¿ will be assigned handling damage as the issue is due to handling of the product or portion of the product during the clinical procedure, upon removal of the product from the packaging, or preparation of the product prior to use.
|