BOSTON SCIENTIFIC CORPORATION CAROTID WALLSTENT; STENT, CAROTID Back to Search Results
Model Number 26606
Device Problem Positioning Failure (1158)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/04/2023
Event Type  malfunction  
Manufacturer Narrative
Device evaluated by manufacturer: carotid wallstent monorail 10.0-31 was received for analysis.A visual and tactile examination identified a complete separation of the outer sheath of the device located approximately 35mm distal of the main t-valve.This type of damage is consistent with excessive force being applied to the device when attempting deployment.No other issues were noted with the delivery system.The device was returned with the stent in the correct position on the delivery system.The investigator was unable to deploy the stent due to a separation of the outer sheath of the device.
 
Event Description
Reportable based on device analysis completed on 10aug2023.A 10.0-31 carotid wallstent was selected for use during a carotid artery stenting procedure to treat the internal carotid artery and common artery.However, during the procedure, the stent could not be deployed.The procedure was completed using another of the same device.No patient complications were reported.However, returned device analysis revealed sheath detached/separated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAROTID WALLSTENT
Type of Device
STENT, CAROTID
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key17633149
MDR Text Key322076689
Report Number2124215-2023-43040
Device Sequence Number1
Product Code NIM
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 08/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/28/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/09/2023
Device Model Number26606
Device Catalogue Number26606
Device Lot Number0025576148
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/02/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/10/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/09/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-