HILL-ROM MEXICO ADVANTA2 RENTAL BED; BED, MANUAL Back to Search Results
Model Number P1190ARENT01
Device Problem Material Split, Cut or Torn (4008)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/03/2023
Event Type  malfunction  
Event Description
Hillrom received a report from a hillrom technician stating the copper wire were exposed.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint (b)(4).
 
Manufacturer Narrative
The hillrom technician found the power cord needed to be replaced.Per the hillrom service manual the advanta¿ 2 should be subject to an effective maintenance program.This will help make sure of a long, operative life for the advanta¿ 2 bed.The preventative maintenance will help to reduce downtime due to excessive wear failures.An annual service of the bed is advised in order to maintain its characteristics and performance.Examine the plug for damage.Make sure the plug is a one-piece molded plug assembly.If it is not, replace the plug cord assembly.Replace any plug cord assembly that shows any of these: discoloration of the plug molding around the plug blades; this could occur if the plug blades have overheated or arced.Any signs of cracking; this could occur if the plug has been bent and straightened to a point past its useful life.Loose fit of the plug blade (the plug blade moves in the molding); this could occur if the molding has overheated or the blades have been bent and straightened to a point past their useful life.Replace the power cord, if damaged.A search of the hillrom maintenance records showed hillrom performed preventative maintenance on this bed on (b)(6) 2022.It is unknown if the facility performed any other preventative maintenance on this bed.The technician will replace the power cord to resolve the reported event.Based on this information, no further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANTA2 RENTAL BED
Type of Device
BED, MANUAL
Manufacturer (Section D)
HILL-ROM MEXICO
ave.del telefono no. 200 col. huinala
apodaca, nuevo leon 66640
MX  66640
Manufacturer Contact
brad wheeler
1069 state route 46 east
batesville, IN 47006
3128199307
MDR Report Key17633988
MDR Text Key322089083
Report Number3006697241-2023-00094
Device Sequence Number1
Product Code FNJ
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 08/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/28/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberP1190ARENT01
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received08/03/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/18/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-