• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AVANOS MEDICAL INC. ON-Q PUMP FIXED FLOW; ELASTOMERIC LFR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AVANOS MEDICAL INC. ON-Q PUMP FIXED FLOW; ELASTOMERIC LFR Back to Search Results
Model Number P400X4D
Device Problem Improper Flow or Infusion (2954)
Patient Problem Insufficient Information (4580)
Event Date 08/01/2023
Event Type  malfunction  
Manufacturer Narrative
The sample is reported to be available but has not yet been received by the manufacturer.A review of the device history record is not possible as no lot number was provided.All information reasonably known as of 24 aug 2023 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the reported event but is relaying the information provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.H3 other text : device not returned.
 
Event Description
Fill volume: unknown.Flow rate: unknown.Procedure: unknown.Cathplace: unknown.Infusion start time: unknown.Infusion stop time: unknown.It was reported, the pump infused into the patient within two days.The pump usually doesn¿t empty for five days.No injury or medical interventions reported.
 
Manufacturer Narrative
Additional information: d1, d4, d9, g4, h6.Correction: d2.The product involved in the report has been returned and the investigation remains in progress at this time.A review of the device history record is in-progress.All information reasonably known as of 10 oct 2023 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the reported event but is relaying the information provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
 
Manufacturer Narrative
The device history record for lot 320222601 was reviewed and the product was produced according to product specifications.The actual complaint product was returned for evaluation.The sample evaluation results revealed that the flow restrictor (glass) was not functioning as intended due to a potential component mix.The incident was confirmed as reported.The root cause was related component mix-up/manufacturing related.All information reasonably known as of 05 jan 2024 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the reported event but is relaying the information provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ON-Q PUMP FIXED FLOW
Type of Device
ELASTOMERIC LFR
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
Manufacturer (Section G)
AVENT S. DE R.L. DE C.V.
ave noruega edificio d-1b
fraccionamiento rubio
tijuana b.c.
Manufacturer Contact
ujjal chakravartty
5405 windward parkway
alpharetta, GA 30004
4704485487
MDR Report Key17636062
MDR Text Key322107519
Report Number2026095-2023-00092
Device Sequence Number1
Product Code MEB
UDI-Device Identifier00193494136765
UDI-Public00193494136765
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063530
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Risk Manager
Type of Report Initial,Followup,Followup
Report Date 01/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/28/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberP400X4D
Device Catalogue NumberN/A
Device Lot Number30222601
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received12/18/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-