Model Number LF2019 |
Device Problems
Energy Output Problem (1431); Noise, Audible (3273)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/04/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
D10: concomitant product: lf2019 - exact dissector lf2019 ligasure 21cm, lot# unknown lf2019 - exact dissector lf2019 ligasure 21cm, lot# unknown vlft10gen- vlft10gen ft series energy platformx1, serial#(b)(6).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, during a thyroid procedure, the first two handpieces did not seal vessel, had hissing noise and steaming.There was activation tone, end tone and energy delivery was noted.The third handpiece would not cut, knife trigger would not move and would not advance.A new handpiece was used with the same generator and it worked fine.There was no patient injury.
|
|
Event Description
|
According to the reporter, during a thyroid procedure, the first two handpieces did not seal vessel, had hissing noise and steaming.There was activation tone, end tone and energy delivery was noted.The third handpiece would not cut, knife trigger would not move and would not advance.A new handpiece was used with the same generator and it worked fine.There was no patient injury.Medtronic's initial evaluation of the incident third device found that the tooth of the blade safety was broken and piece was not returned.
|
|
Manufacturer Narrative
|
Additional information: h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the ligasure produced only a partial seal of the vessel and the device was had an unusual sound.The reported issues could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, during a thyroid procedure, the first two handpieces did not seal vessel, had hissing noise and steaming.There was activation tone, end tone and energy delivery was noted.The third handpiece would not cut, knife trigger would not move and would not advance.A new handpiece was used with the same generator and it worked fine.There was no detached piece fell on patient's cavity.There was no patient injury.Medtronic's initial evaluation of the incident third device found that the tooth of the blade safety was broken and piece was not returned.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|