• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COMPANION MEDICAL INC INPEN MMT-105NNBLNA NOVO NORDISK BLUE; SYRINGE, PISTON

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COMPANION MEDICAL INC INPEN MMT-105NNBLNA NOVO NORDISK BLUE; SYRINGE, PISTON Back to Search Results
Model Number MMT-105NNBLNA
Device Problems Break (1069); Fluid/Blood Leak (1250); Patient Device Interaction Problem (4001); Priming Problem (4040)
Patient Problem Hyperglycemia (1905)
Event Date 08/11/2023
Event Type  malfunction  
Event Description
Information received by medtronic indicated that the customer reported the insulin squirting out of the inpen after injecting insulin.Troubleshooting was performed and drips of insulin came out after an injection was done.The button pressed down the inpen and was pulled out, and altitude and temperature changed.No harm requiring medical intervention was reported.The inpen was returned for analysis.
 
Manufacturer Narrative
Per visual inspection: no physical damage to cartridge holder or inpen front or back shell was noted.The inpen paired to the commercial app.The leadscrew was received 1/2 it's travel.Inpen passed baseline and wireless functionality.App logbook displayed: 15u, 15u, 15u, 15u, 15u, 15u, 15u, 15u, 15u, 15u.Performed leak test and no fluid leaks were noted outside the fluid pathway.Inpen passed front cap investigation.In conclusion: no fluid leaks were noted on inpen.Inpen working as design.Insulin leaks not confirmed.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INPEN MMT-105NNBLNA NOVO NORDISK BLUE
Type of Device
SYRINGE, PISTON
Manufacturer (Section D)
COMPANION MEDICAL INC
1223 world trade drive, 100
san diego CA 92128
Manufacturer (Section G)
COMPANION MEDICAL INC
1223 world trade drive, 100
san diego CA 92128
Manufacturer Contact
justin ellis
1223 world trade drive, 100
san diego, CA 92128
7635265677
MDR Report Key17640023
MDR Text Key322152783
Report Number3012822846-2023-01515
Device Sequence Number1
Product Code FMF
UDI-Device Identifier000010862088000344
UDI-Public(01)000010862088000344(17)240428(10)B1098
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K201337
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 08/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/28/2024
Device Model NumberMMT-105NNBLNA
Device Catalogue NumberMMT-105NNBLNA
Device Lot NumberB1098
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/22/2023
Initial Date Manufacturer Received 08/24/2023
Initial Date FDA Received08/29/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
Patient Age4 YR
Patient SexMale
-
-