HILL-ROM BATESVILLE TOTALCARE FRAME; BED, AC-POWERED ADJUSTABLE HOSPITAL Back to Search Results
Model Number P1900Q007328
Device Problem Activation Problem (4042)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/17/2023
Event Type  malfunction  
Event Description
Hillrom received a report from a hillrom technician stating the bed had a stuck cpr valve and the cpr function was inoperative.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint # (b)(4).
 
Manufacturer Narrative
The hillrom technician found the cpr valve needed to be replaced.Per the hillrom service manual, the totalcare bed system requires an effective maintenance program.We recommend that you perform a semi-annual preventative maintenance.Test the cpr release for correct operation and reset of the head cylinder.A search of the hillrom maintenance records did not show hillrom performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The technician replaced the cpr valve to resolve the reported event.Based on this information, no further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOTALCARE FRAME
Type of Device
BED, AC-POWERED ADJUSTABLE HOSPITAL
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
kayla miller
1069 state route 46 east
batesville, IN 47006
8129310130
MDR Report Key17641898
MDR Text Key322169705
Report Number1824206-2023-00984
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K962942
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 08/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberP1900Q007328
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/17/2023
Initial Date FDA Received08/29/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/03/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-