Catalog Number SPD2-070-320
Device Problems Difficult to Remove (1528); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/01/2023
Event Type  Injury  
Event Description
A physician was using a spider fx during treatment of a 120mm calcified lesion with 60% stenosis in the distal region of the superficial femoral artery and popliteal artery.The artery was 6mm in diameter with moderate calcification and mild tortuosity.A 7fr sheath was used.The vessel was not pre-dilated, the vessel was post-dilated.The ifu was followed.It was reported during g withdrawal of 7 mm filter; the device became stuck inside the patient due to a large piece of calcium contained in the filter. the filter was attempted to be withdrawn into the imaging catheter <(>&<)> then the 7fr sheath.The filter was unable to be withdrawn into either the 5fr catheter or 7 fr sheath.Surgeon pulled at filter repeatedly and the wire then became detached leading to filter basket remaining in the common femoral artery. a cutdown of the common femoral artery was performed and the filter was successfully removed.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPIDER FX
Type of Device
TEMPORARY CAROTID CATHETER FOR EMBOLIC CAPTURE
Manufacturer (Section D)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key17641920
MDR Text Key322169808
Report Number2183870-2023-00310
Device Sequence Number1
Product Code NTE
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K111010
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 08/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberSPD2-070-320
Device Lot NumberB481629
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/07/2023
Initial Date FDA Received08/29/2023
Date Device Manufactured11/22/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient SexMale
-
-