• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. LIGACLIP*ENDO ROTATING MCA; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. LIGACLIP*ENDO ROTATING MCA; CLIP, IMPLANTABLE Back to Search Results
Catalog Number ER320
Device Problems Failure to Fire (2610); Difficult to Open or Close (2921)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/12/2023
Event Type  malfunction  
Event Description
It was reported that during a robot assisted rectum procedure, the trigger was stiff at 3rd firing.The device was inserted from the assisted port.The device locked out when the trigger was grasped outside patient.Manual override lever was used, but the jaws did not open.Hem-o-lok was used to complete the case.There were no adverse consequences to the patient.No further information is available.
 
Manufacturer Narrative
(b)(4).Date sent: 8/29/2023.D4: batch # a9c22x.Investigation summary: the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the er320 device was returned with the jaws closed and with no apparent damage.In an attempt to replicate the reported incident, the device was tested for functionality.Upon testing, the trigger could not be activated due to be jammed.The device was disassembled to evaluate the condition of the internal components and the trigger was found bent, not allowing the clips to fed into the jaws.In addition, 1 conforming clip that was in the jaws and 16 clips were found remaining inside the clip track.The damaged on the device could be that the internal ribs were being gauged by the feeder link, causing the trigger to experienced additional force that could interfere with the firing of the device.As part of ethicon¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch and lot number, and no non-conformances were identified.
 
Manufacturer Narrative
(b)(4).Date sent: 2/14/2024.Additional analysis was completed on the device: "the cracks and crazing in the twist area of the trigger indicate a mechanical failure.The cracks and crazing in the slot region of the trigger indicate mechanical damage.There is crazing and damage on other parts of the trigger.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
3035526892
MDR Report Key17642155
MDR Text Key322175071
Report Number3005075853-2023-06202
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10705036012580
UDI-Public10705036012580
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K864102
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberER320
Device Lot Number278C10
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/28/2023
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 08/02/2023
Initial Date FDA Received08/29/2023
Supplement Dates Manufacturer Received01/31/2024
Supplement Dates FDA Received02/14/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/13/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-