It was reported that during a robot assisted rectum procedure, the trigger was stiff at 3rd firing.The device was inserted from the assisted port.The device locked out when the trigger was grasped outside patient.Manual override lever was used, but the jaws did not open.Hem-o-lok was used to complete the case.There were no adverse consequences to the patient.No further information is available.
|
(b)(4).Date sent: 8/29/2023.D4: batch # a9c22x.Investigation summary: the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the er320 device was returned with the jaws closed and with no apparent damage.In an attempt to replicate the reported incident, the device was tested for functionality.Upon testing, the trigger could not be activated due to be jammed.The device was disassembled to evaluate the condition of the internal components and the trigger was found bent, not allowing the clips to fed into the jaws.In addition, 1 conforming clip that was in the jaws and 16 clips were found remaining inside the clip track.The damaged on the device could be that the internal ribs were being gauged by the feeder link, causing the trigger to experienced additional force that could interfere with the firing of the device.As part of ethicon¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch and lot number, and no non-conformances were identified.
|