ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. SAFESET¿ - TRANSPAC® IT W/03 ML RESERVOIR AND NEEDLELESS VALVE, RED STRIPE TUBIN; TRANSDUCER, BLOOD-PRESSURE, EXTRAVASCULAR
|
Back to Search Results |
|
Catalog Number 011-46103-90 |
Device Problem
Separation Problem (4043)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/01/2023 |
Event Type
malfunction
|
Event Description
|
The event occurred on an unspecified date and involved a safeset transpac® it w/03 ml reservoir and needleless valve, red stripe tubing, with velcro arm strap, patient mount device where it was reported the glue that sticks the barrel to the arterial line transducer came undone and the patient lost blood.There was patient involvement and unknown patient harm.
|
|
Manufacturer Narrative
|
The device is available for evaluation, however has not been received.
|
|
Manufacturer Narrative
|
Reveived one used partial.List #011-46103-90, safeset¿ - transpac® it w/03 ml reservoir and needleless valve, red stripe tubing, with velcro arm strap, patient mount; lot #13534687.One used.Unknown manufacturer, unknown tubing; lot #unknown.The reported complaint of separation was confirmed.During visual inspection, only the rotating male luer connecting to the unknown tubing was received for evaluation.The separated tubing from the rotating male luer was not received for evaluation.It is unknown how and when the tubing went separated from the rotating male luer.Without the return of the entire set, which was connected to the rotating male luer, a probable cause could not be determined.The lot history was reviewed, no nonconformities were identified that may have contributed to the reported complaint.D9 device received on 8/23/2023.
|
|
Search Alerts/Recalls
|
|
|