AVANOS MEDICAL INC. CORTRAK 2 NG/NI FEEDING TUBE, ELECTRO STYLET, ENFIT; DH CORTRAK DISPOSABLES
|
Back to Search Results |
|
Model Number 40-9551TRAK2 |
Device Problem
Break (1069)
|
Patient Problem
Insufficient Information (4580)
|
Event Type
malfunction
|
Event Description
|
It was reported there was a "severed corpak" ng (nasogastric) tube noted during use.No additional information was provided.
|
|
Manufacturer Narrative
|
The actual device is reported to be available but has not been returned for evaluation.A review of the device history record is not possible as no lot number was provided.All information reasonably known as of 28-aug-2023 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
|
|
Manufacturer Narrative
|
The product involved in the report has been returned and is being processed for evaluation.All information reasonably known as of 29-sep-2023 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21 cfr 803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
|
|
Manufacturer Narrative
|
The device was evaluated.The tubing appeared to have expanded to form a balloon shape which burst radially causing a separation of the tube into two pieces.When manually pressing the tubing, there was thin layer of the material noticed on the ballooned portion of the tube.The breaking point exhibited to have an unknown dried-hardened material residue which wasn't able to be removed.Root cause could not be determined.All information reasonably known as of 27-oct-2023 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
|
|
Search Alerts/Recalls
|
|
|