ABBOTT QUADRA ASSURA MP ICD; NO MATCH Back to Search Results
Model Number CD3371-40QC
Device Problems Environmental Compatibility Problem (2929); Inappropriate or Unexpected Reset (2959)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/09/2023
Event Type  malfunction  
Event Description
Following an mri procedure, the device was observed to be in backup mode due to not programming the device to mri mode resulting in a loss of defibrillation therapy.No therapy was required while the device was in backup mode.Abbott technical support was contacted and the device was reprogrammed to resolve the event.The patient was in stable condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA MP ICD
Type of Device
NO MATCH
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17646379
MDR Text Key322222709
Report Number2017865-2023-40602
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2020
Device Model NumberCD3371-40QC
Device Lot NumberP000059274
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/09/2023
Initial Date FDA Received08/29/2023
Supplement Dates Manufacturer Received11/07/2023
Supplement Dates FDA Received11/07/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/30/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexMale
-
-