Model Number HL-90 |
Device Problems
Device Alarm System (1012); Alarm Not Visible (1022)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
It was reported that the customer conducted an alarm test where the green light turns on, but the other lights did not.The disposable light not turning on.Low water light not working.Overtemperature never alarms when pressed on side.No adverse effects have been reported.
|
|
Manufacturer Narrative
|
Other, other text: b3: date of event is unknown no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Event Description
|
Additional information received via email.The customer stated the problem was detected during the incoming inspection by their clinical equipment department.No patient harm and medical intervention involved.
|
|
Manufacturer Narrative
|
Device available for evaluation; device evaluated by manufacturer and h6.Evaluation codes: email is: regulatory.Responses@icumed.Com.One device was received in good condition.No physical observations were noted.Per functional testing, the light emitting diodes (led)s did not light up.When they were pressed, they did light up, but wouldn't stay on.The complaint was confirmed.The root cause was a faulty printed circuit board (pcb).It was unknown what caused it to become faulty.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.Replaced pcb.Performed preventative maintenance (pm), changed o-rings and reservoir gasket and calibrated.The device passed all functional and delivery tests.
|
|