SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX TUBES BLUSELECT SUCTIONAID; TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
|
Back to Search Results |
|
Model Number 101/810/080 |
Device Problem
Material Puncture/Hole (1504)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/01/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that the cuff tore during use (one week after intubation).No patient injury or clinical affects was reported.
|
|
Manufacturer Narrative
|
Other, other text: b3: month and year of event have been provided, day is unknown.D4: lot number, expiration date and h4: device manufacture date is unknown, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Manufacturer Narrative
|
Other text: h3, h6: event methods, evaluation, and conclusion codes: updated.One device was received for investigation.The reported issue was confirmed during visual inspection, which demonstrated that the inflation line had been detached from the pilot balloon.However, no root cause could be identified for this condition.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.Complaint information will continue to be monitored and actions assigned as required.
|
|