Model Number P3200K000123
Device Problem Activation Problem (4042)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/04/2023
Event Type  malfunction  
Manufacturer Narrative
The hillrom technician found the cpr cable harness connection needed to be reseated.Examine the condition of the two cpr release handles, cpr cable, and cpr mechanism on the head motor.Make sure the screws are installed and fully tightened.Operate the head section to the high position, then engage one of the cpr releases.Make sure the head section lowers.Adjust the cpr cable as necessary.Do the same tests on the other side of the bed.Release the cpr handles and make sure the mechanism locks correctly when you operate the head up control for a few seconds.The head section must rise.Replace the head section motor in the event of a malfunction or the cpr cable if broken.A search of the hillrom maintenance records did not show hillrom performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The technician reseated the cpr cable harness connection to resolve the reported event.Based on this information, no further action is required.
 
Event Description
Hillrom received a report from a hillrom technician stating the upper control board leds were showing a cpr switch failure error and cpr was inoperative.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint #(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSACARE FRAME
Type of Device
BED, AC-POWERED ADJUSTABLE HOSPITAL
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
brad wheeler
1069 state route 46 east
batesville, IN 47006
3128199307
MDR Report Key17650444
MDR Text Key322281351
Report Number1824206-2023-00995
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 08/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/30/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberP3200K000123
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received08/04/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/26/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-