The hillrom technician found the cpr cable harness connection needed to be reseated.Examine the condition of the two cpr release handles, cpr cable, and cpr mechanism on the head motor.Make sure the screws are installed and fully tightened.Operate the head section to the high position, then engage one of the cpr releases.Make sure the head section lowers.Adjust the cpr cable as necessary.Do the same tests on the other side of the bed.Release the cpr handles and make sure the mechanism locks correctly when you operate the head up control for a few seconds.The head section must rise.Replace the head section motor in the event of a malfunction or the cpr cable if broken.A search of the hillrom maintenance records did not show hillrom performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The technician reseated the cpr cable harness connection to resolve the reported event.Based on this information, no further action is required.
|