A caller reported experiencing a skin reaction while wearing the adc device.The customer experienced symptoms described as itching, redness, and allergy at the sensor insertion site.The customer had contact with a healthcare professional who applied an unspecified medication (spray) at the sensor insertion site for treatment.No further information was provided.There was no report of death or permanent injury associated with this event.
|
Repeated attempts by adc to retrieve the product were unsuccessful and/or the customer discarded the product.The most probable root causes associated with this failure mode are disconnected, faulty or damaged components, or misuse.However, mitigations are in place to reduce and prevent such issues.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.At this time product has not yet been returned for this complaint.An extended investigation has been performed for the reported complaint.There was no indication that the product did not meet specification.Dose audit reports were reviewed and demonstrates the continued effectiveness of the established sterilization process for libre sensor products.Environmental monitoring reports were reviewed, including bioburden and endotoxin testing, and demonstrated that all monitoring processes continue to meet adc minimum requirements for product quality.The reported complaint is related to skin irritation or an allergic reaction to the patch adhesive of the freestyle libre sensor.The dhrs (device history review) for libre sensor kits were reviewed and the dhrs showed the libre sensor kits passed all tests prior to release.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
|
A caller reported experiencing a skin reaction while wearing the adc device.The customer experienced symptoms described as itching, redness, and allergy at the sensor insertion site.The customer had contact with a healthcare professional who applied an unspecified medication (spray) at the sensor insertion site for treatment.No further information was provided.There was no report of death or permanent injury associated with this event.
|