• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. ZIMMER AIR DERMATOME; DERMATOME HANDPIECE, PNEUMATIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. ZIMMER AIR DERMATOME; DERMATOME HANDPIECE, PNEUMATIC Back to Search Results
Model Number N/A
Device Problems Power Problem (3010); Appropriate Term/Code Not Available (3191); Intermittent Loss of Power (4016)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/08/2023
Event Type  malfunction  
Manufacturer Narrative
This complaint is recorded by zimmer biomet under (b)(4).A follow up/ final report will be submitted once investigation is complete.G2: foreign country - sweden.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that during surgery on (b)(6) 2023 slower phase when on and then stopped working for a few seconds then started working again at a slower phase.There was no harm or delay reported.Due diligence is complete and there is no additional information till date.
 
Manufacturer Narrative
Review of the most recent repair record determined the motor speed was unstable and there was corrosion on the needle bearing e-ring and reciprocation arm.The motor, needle bearing, e-ring, and reciprocation arm were replaced and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
There is no additional event information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIMMER AIR DERMATOME
Type of Device
DERMATOME HANDPIECE, PNEUMATIC
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key17650747
MDR Text Key322286101
Report Number0001526350-2023-01019
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
Reporter Country CodeSW
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/30/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00880100100
Device Lot Number61480980
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/16/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/14/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-