Updated fields: d4, d9, g3, g6, h2, h3, h4, h10.The mayfield composite series skull clamp (a3059) was returned for evaluation: device history record (dhr) - the dhr was reviewed and shows no abnormalities related to the reported failure.Failure analysis - investigation of the returned unit was unable to duplicate any slippage.The unit does have up and down movement in the lock, but when the unit was properly positioned and put under pressure, it did not slip.The disk ratchet and retainer were replaced due to being worn, and the 80lb torque knob is the older style that was replaced with the current revision.All other worn components were replaced with new parts, and general maintenance and cleaning were performed.Root cause - the complaint is not confirmed via inspection of the unit.The probable cause of the present movement is routine wear of the unit.No further investigation is required based on the acceptability of risk and no adverse trends were identified.This will be monitored and trended going forward.At present, we consider this complaint to be closed.
|