Catalog Number SGC0701
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/09/2023
Event Type  malfunction  
Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event Description
It was reported this was a mitraclip procedure to treat mitral regurgitation (mr) with a grade of 4.Prior to the removal of the dilator, the steerable guide catheter (sgc) hemostasis valve flush port broke.The sgc was replaced and the procedure continued without issue.The mr grade reduced to 1.There were no adverse patient effects and no clinically significant delay in the procedure.
 
Manufacturer Narrative
In this case the returned device analysis confirmed the reported flush port crack.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar complaints reported from this lot.Based on available information and without the device to analyze, the reported flush port crack was determined to be due to a potential product issue.Abbott will continue to trend the performance of these devices.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP¿ G4 STEERABLE GUIDE CATHETER (MDR)
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key17669417
MDR Text Key322481798
Report Number2135147-2023-03818
Device Sequence Number1
Product Code DRA
UDI-Device Identifier08717648231025
UDI-Public08717648231025
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K190167
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/25/2024
Device Catalogue NumberSGC0701
Device Lot Number30327R1041
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/09/2023
Initial Date FDA Received09/01/2023
Supplement Dates Manufacturer Received10/25/2023
Supplement Dates FDA Received11/16/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/27/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
STEERABLE GUIDE CATHETER.
-
-